Planmeca – Panoramic Dental X-Ray Unit – Viso Series

0,00 

Viso G7

Classification Medical Device Directive: 93/42/EEC (Class IIb)
RoHS: 2011/65/EU
IEC60601-1: Class I, type B
CISPR: Class B
IP Classification: IPX0
Generator Constant potential, microprocessor controlled, resonance mode, operating
frequency 80 -160 kHz, Power Factor Corrector, complies with the standard IEC
60601-2-7: 1998
X-ray tube D-059SBR or SXR 130-10-0.5 SC
Focal spot size 0.5×0.5mm, according to IEC 60336: 2005
Total filtration 3D: Total 2.5 mm Al + 0.2/0.5 mm Cu
Pan: Total 2.5 mm Al
Tube housing front cover quality equivalent filtration: 0.3 mm Al @ 70 kV / HVL
2.6 mm AL
Anode voltage 3D: 60 – 120 kV ± 5%
Pan: 60– 84 kV ± 5%
Anode current D-059SBR: 1 – 14 mA (3D: 1 – 12.5 mA)
SXR 130-10-0.5 SC: 1 – 16 mA
(mAs steps with R20 series: 2.0, 2.2, 2.5, 2.8, 3.2, 3.6, 4.0, 4.5, 5.0, 5.6, 6.3, 7.1,
8.0, 9.0, 10, 11, 12.5, 14, 16, 18, 20, 22, 25, 28, 32, 36, 40, 45, 50, 56, 63, 71, 80,
90, 100, 110, 125, 140, 160, 180, 200, 220, 250)
Exposure time 3D: 1 – 36 s effective, pulsed
Pan: 2.5 – 15.6 s
Scan angle 210° / 360°
SID 700 mm
Focus to skin distance min. 150 mm
Magnification 3D: 1.40x – 1.71x
Pan: 1.40x
Line voltage 100 – 220 V~/ 50 or 60 Hz, Power Factor Corrector
230 – 240 V~/ 50 Hz, Power Factor Corrector
Line current 8 – 17 A
Input power Stand by: 150 VA
Exposure 1800 W
Max continuous heat
dissipation
250 W
Fuses One user replaceable fuse
16 A FF H 500 V or 8 A FF H 500 V
Type: 195100 ELU

Out of stock

SKU: 09739 Category:

Description

Viso G7

Classification Medical Device Directive: 93/42/EEC (Class IIb)
RoHS: 2011/65/EU
IEC60601-1: Class I, type B
CISPR: Class B
IP Classification: IPX0
Generator Constant potential, microprocessor controlled, resonance mode, operating
frequency 80 -160 kHz, Power Factor Corrector, complies with the standard IEC
60601-2-7: 1998
X-ray tube D-059SBR or SXR 130-10-0.5 SC
Focal spot size 0.5×0.5mm, according to IEC 60336: 2005
Total filtration 3D: Total 2.5 mm Al + 0.2/0.5 mm Cu
Pan: Total 2.5 mm Al
Tube housing front cover quality equivalent filtration: 0.3 mm Al @ 70 kV / HVL
2.6 mm AL
Anode voltage 3D: 60 – 120 kV ± 5%
Pan: 60– 84 kV ± 5%
Anode current D-059SBR: 1 – 14 mA (3D: 1 – 12.5 mA)
SXR 130-10-0.5 SC: 1 – 16 mA
(mAs steps with R20 series: 2.0, 2.2, 2.5, 2.8, 3.2, 3.6, 4.0, 4.5, 5.0, 5.6, 6.3, 7.1,
8.0, 9.0, 10, 11, 12.5, 14, 16, 18, 20, 22, 25, 28, 32, 36, 40, 45, 50, 56, 63, 71, 80,
90, 100, 110, 125, 140, 160, 180, 200, 220, 250)
Exposure time 3D: 1 – 36 s effective, pulsed
Pan: 2.5 – 15.6 s
Scan angle 210° / 360°
SID 700 mm
Focus to skin distance min. 150 mm
Magnification 3D: 1.40x – 1.71x
Pan: 1.40x
Line voltage 100 – 220 V~/ 50 or 60 Hz, Power Factor Corrector
230 – 240 V~/ 50 Hz, Power Factor Corrector
Line current 8 – 17 A
Input power Stand by: 150 VA
Exposure 1800 W
Max continuous heat
dissipation
250 W
Fuses One user replaceable fuse
16 A FF H 500 V or 8 A FF H 500 V
Type: 195100 ELU

Additional information

Type

Producer